
14 RS232 ASCII Interface Protocol
jins85e1 2.fm Interface Description(1101)
Command
word1
Command
word2
Command
word3
Parameter Statement Read/
Set
Example Response
*CONFig :UNIT :FlowUnit unit of the carrier gas flow –
<unit> unit = sccm or l/s S
R
*CONF:UNIT:FU sccm
*CONF:UNIT:FU?
OK
sccm
:LR leak rate unit –
<unit> unit = SCCM; MBAR*L/S; PA*M3/S;
ATM*CC/S; TORR*L/S
S
R
*CONF:UNIT:LR TORR*L/S
*CONF:UNIT:LR?
OK
Torr*l/s
:Pressure pressure unit –
<unit> unit = MBAR; PA; ATM; TORR S
R
*CONF:UNIT:P MBAR
*CONF:INIT:P?
OK
mbar
:TLUnit unit of the external test leak –
<unit> unit = SCCM; MBAR*L/S; PA*M3/S;
ATM*CC/S; TORR*L/S
:VolUnit unit of the free accumulation volume –
<unit> unit = LITER; CUBICIN; CUBICFT; CCM S
R
*CONF:UNIT:VU CCM
*CONF:UNIT:VU?
OK
CCM
*END interrupt or stop measurement and switch to
READY
S
*GROSS switch to gross mode (only in cont. mode) –
<b> b = on or off S
R
*GROSS ON
*GROSS?
OK
ON
*HOUR :DATE read / write date
<value> value = dd.mm.yyyy R/S *HOUR:DATE? 03.02.2009
:DEVICE total operation time of T-Guard R *HOUR:DEVICE? 2039.1h
:SINCE time after switch on (12:23; hh:mm) R *HOUR:SINCE? 00:40 hh:mm
:TIME read/write time (real clock time) –
Commenti su questo manuale