6 RS232 ASCII Interface Protocol
jins85e1 2.fm Interface Description(1101)
2.3 List of ASCII Commands
Command
word1
Command
word2
Command
word3
Parameter Statement Read
Set
Example Response
*CAL only external calibration possible
–
:START S*CAL:START OK
:LEAKrate <value> leak rate of test leak R
S
*CAL:LEAK?
*CAL:LEAK 2.4E-5
1.20E-5
OK
:UNIT unit of test leak
–
MBAR*L/S R
S
*CAL:UNIT?
*CAL:UNIT MBAR*L/S
mbar*l/s
OK
PA*M3/S R/S *CAL:UNIT PA*M3/S OK
SCCM R/S *CAL:UNIT SCCM OK
TORR*L/S R/S *CAL:UNIT TORR*L/S OK
:QUIT acknowledge (confirm) after calibration S *CAL:QUIT OK
:ESC escape after calibration S *CAL:ESC OK
:FACtor calibration factor (values only available after a
finished calibration)
–
:NEW actual calibration factor
(only available directly after a calibration)
R *CAL:FAC:NEW? 0.576
:OLD last calibration factor R *CAL:FAC:OLD? 0.974
:PRESsure pressure at the Wise Technology
TM
Sensor
–
:NEW actual calibration R *CAL:PRES:NEW? 271
:OLD last calibration R *CAL:PRES:OLD? 263
:StartProof starts proof function S