14 Cursor down to START. Press F3 INSERT. Start ### will appear.
15 Press F1 IF/THEN.Cursor to ### behind Start and enter 1(the Material to start).
Press ENTER.
16 Press F4 SAVE & EXIT to return to the Logic page. At this point the Logic
Statement is entered into memory.
17 To return to the OPERATE display, press Menu.
This example allows starting Material 1 only if it is in Ready or Idle but not when
Material 1 is in Stop.